Column |
Column Title |
Description |
A |
Institution |
Anonymous version of institution name. |
B |
Machine |
Anonymous version of machine ID. |
C |
Acquisition |
Date and time that the first image in the data set was captured (delivery time). |
D |
Analysis |
Date and time that data was analyzed. |
E |
Procedure |
Procedure (test) used to process data |
F |
Beam Name |
Common name of RTPLAN beam. |
G |
Baseline Designation |
explicit: Designated by user as a baseline. implicit: Used as baseline when an explicit one is not defined. If blank, then it is not used as a baseline. |
H |
Wedge CU |
Average CU of pixels in the center of the wedge image. |
I |
Background Beam Name |
Common name of RTPLAN for background beam. |
J |
Background CU |
For background beam image, average CU of pixels in the center. |
K |
Percent of Background |
Wedge CU / Background CU * 100 |
L |
Baseline Wedge CU |
For baseline wedge image, average CU of pixels in the center. |
M |
Baseline Background CU |
For background baseline wedge image, average CU of pixels in the center. |
N |
Baseline Percent of Background |
For background, wedge CU / background CU * 100 |
O |
Machine Type |
Type of machine |
P |
Collimator |
Type of collimator |
Q |
EPID |
Type of EPID |
R |
URL |
Link to main report. |
S |
outputPK |
Internal database reference to synthetic public key of output. |
T |
inputPK |
Internal database reference to synthetic public key of input. |
U |
Uploaded By |
Anonymized version of user ID. |
V |
Pixel Rows |
Number of rows in the images. DICOM 0028,0010 |
W |
Pixel Columns |
Number of columns in the images. DICOM 0028,0011 |
X |
Pixel size X |
Image Plane Pixel Spacing. DICOM 3002,0011 |
Y |
Pixel size Y |
Image Plane Pixel Spacing. DICOM 3002,0011 |
Z |
b value for converting pixel values to CU: CU = m*PixelValue + b |
Rescale Intercept. DICOM 0028,1052 |
AA |
m value for converting pixel values to CU: CU = m*PixelValue + b |
Rescale Slope. DICOM 0028,1053 |
AB |
KVP |
Peak kilo voltage output of the X-Ray generator used. DICOM 0018,0060 |
AC |
Exposure Time msec |
Exposure time in mseync. DICOM 0016,0004 |
AD |
Meterset Exposure |
Treatment machine Meterset duration over which image has been acquired, specified in Monitor units (MU) or minutes as defined by Primary Dosimeter Unit (300A,00B3). Required if Value 3 of Image Type (0008,0008) is PORTAL DICOM 3002,0032 |
AE |
Start Cumulative Meterset Wt |
Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition starts. DICOM 300C,0008 |
AF |
X-Ray Origin X |
Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D |
AG |
X-Ray Origin Y |
Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D |
AH |
X-Ray Origin Z |
Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D |
AI |
Source to Image |
Distance from radiation machine source to image plane (in mm) along radiation beam axis. DICOM 3002,0026 |
AJ |
Source to Isocenter |
Radiation source to Gantry rotation axis distance of radiation machine used in acquiring or computing image (mm). DICOM 3002,0022 |
AK |
Gantry Angle |
Treatment machine gantry angle, i.e., orientation of IEC GANTRY coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees). DICOM 300A,011E |
AL |
Collimator Angle |
Treatment machine beam limiting device (collimator) angle, i.e., orientation of IEC BEAM LIMITING DEVICE coordinate system with respect to IEC GANTRY coordinate system (degrees). DICOM 300A,0120 |
AM |
End Cumulative Meterset Wt |
Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition ends. DICOM 300C,0009 |
AN |
Beam Number |
Uniquely identifies the corresponding N-segment treatment beam specified by Beam Number (300A,00C0) within Beam Sequence (300A,00B0) in RT Beams Module within the RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or the Ion Beam Sequence (300A,03A2) in the RT Ion Beams Module within the RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). DICOM 300C,0006 |
AO |
Operator |
Anonymized name(s) of the operator(s) supporting the Series. DICOM 0008,1070 |
AP |
Software Version |
Manufacturer's designation of software version of the equipment that produced the sources. DICOM 0018,1020 |
AQ |
SOPInstanceUID |
Uniquely identifies the SOP Instance. DICOM 0008,0018 |
AR |
SeriesInstanceUID |
Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object instance(s). DICOM 0020,000E |
AS |
PatientID |
Primary identifier for the Patient. DICOM 0010,0020 |
AT |
PatientName |
Patient's full name DICOM 0010,0010 |
AU |
Background Pixel Rows |
Background Number of rows in the images. DICOM 0028,0010 |
AV |
Background Pixel Columns |
Background Number of columns in the images. DICOM 0028,0011 |
AW |
Background Pixel size X |
Background Image Plane Pixel Spacing. DICOM 3002,0011 |
AX |
Background Pixel size Y |
Background Image Plane Pixel Spacing. DICOM 3002,0011 |
AY |
Background b value for converting pixel values to CU: CU = m*PixelValue + b |
Background Rescale Intercept. DICOM 0028,1052 |
AZ |
Background m value for converting pixel values to CU: CU = m*PixelValue + b |
Background Rescale Slope. DICOM 0028,1053 |
BA |
Background KVP |
Background Peak kilo voltage output of the X-Ray generator used. DICOM 0018,0060 |
BB |
Background Exposure Time msec |
Background Exposure time in mseync. DICOM 0016,0004 |
BC |
Background Meterset Exposure |
Background Treatment machine Meterset duration over which image has been acquired, specified in Monitor units (MU) or minutes as defined by Primary Dosimeter Unit (300A,00B3). Required if Value 3 of Image Type (0008,0008) is PORTAL DICOM 3002,0032 |
BD |
Background Start Cumulative Meterset Wt |
Background Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition starts. DICOM 300C,0008 |
BE |
Background X-Ray Origin X |
Background Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D |
BF |
Background X-Ray Origin Y |
Background Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D |
BG |
Background X-Ray Origin Z |
Background Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D |
BH |
Background Source to Image |
Background Distance from radiation machine source to image plane (in mm) along radiation beam axis. DICOM 3002,0026 |
BI |
Background Source to Isocenter |
Background Radiation source to Gantry rotation axis distance of radiation machine used in acquiring or computing image (mm). DICOM 3002,0022 |
BJ |
Background Gantry Angle |
Background Treatment machine gantry angle, i.e., orientation of IEC GANTRY coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees). DICOM 300A,011E |
BK |
Background Collimator Angle |
Background Treatment machine beam limiting device (collimator) angle, i.e., orientation of IEC BEAM LIMITING DEVICE coordinate system with respect to IEC GANTRY coordinate system (degrees). DICOM 300A,0120 |
BL |
Background End Cumulative Meterset Wt |
Background Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition ends. DICOM 300C,0009 |
BM |
Background Beam Number |
Background Uniquely identifies the corresponding N-segment treatment beam specified by Beam Number (300A,00C0) within Beam Sequence (300A,00B0) in RT Beams Module within the RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or the Ion Beam Sequence (300A,03A2) in the RT Ion Beams Module within the RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). DICOM 300C,0006 |
BN |
Background Operator |
Background Anonymized name(s) of the operator(s) supporting the Series. DICOM 0008,1070 |
BO |
Background Software Version |
Background Manufacturer's designation of software version of the equipment that produced the sources. DICOM 0018,1020 |
BP |
Background SOPInstanceUID |
Background Uniquely identifies the SOP Instance. DICOM 0008,0018 |
BQ |
Background SeriesInstanceUID |
Background Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object instance(s). DICOM 0020,000E |
BR |
Background PatientID |
Background Primary identifier for the Patient. DICOM 0010,0020 |
BS |
Background PatientName |
Background Patient's full name DICOM 0010,0010 |
BT |
Note |