Definitions for VMAT T2-DG-RS CSV columns.

Column Column Title Description
A Institution Anonymous version of institution name.
B Machine Anonymous version of machine ID.
C Acquisition Date and time that the first image in the data set was captured (delivery time).
D Analysis Date and time that data was analyzed.
E Procedure Procedure (test) used to process data
F Beam Name MLC Name of RTPLAN MLC beam.
G Beam Name Open Name of RTPLAN Open beam.
H R LS -5.1 cm Avg CU of T3MLCSpeed for AOI centered at X = -5.1 cm
I R LS -3.1 cm Avg CU of T3MLCSpeed for AOI centered at X = -3.1 cm
J R LS -1.1 cm Avg CU of T3MLCSpeed for AOI centered at X = -1.1 cm
K R LS 0.9 cm Avg CU of T3MLCSpeed for AOI centered at X = 0.9 cm
L R LS 2.9 cm Avg CU of T3MLCSpeed for AOI centered at X = 2.9 cm
M R LS 4.9 cm Avg CU of T3MLCSpeed for AOI centered at X = 4.9 cm
N R LS 6.9 cm Avg CU of T3MLCSpeed for AOI centered at X = 6.9 cm
O R Open -5.1 cm Avg CU of T3 Open for AOI centered at X = -5.1 cm
P R Open -3.1 cm Avg CU of T3 Open for AOI centered at X = -3.1 cm
Q R Open -1.1 cm Avg CU of T3 Open for AOI centered at X = -1.1 cm
R R Open 0.9 cm Avg CU of T3 Open for AOI centered at X = 0.9 cm
S R Open 2.9 cm Avg CU of T3 Open for AOI centered at X = 2.9 cm
T R Open 4.9 cm Avg CU of T3 Open for AOI centered at X = 4.9 cm
U R Open 6.9 cm Avg CU of T3 Open for AOI centered at X = 6.9 cm
V R Corr -5.1 cm 100 * LS / Open for AOI centered at X = -5.1 cm
W R Corr -3.1 cm 100 * LS / Open for AOI centered at X = -3.1 cm
X R Corr -1.1 cm 100 * LS / Open for AOI centered at X = -1.1 cm
Y R Corr 0.9 cm 100 * LS / Open for AOI centered at X = 0.9 cm
Z R Corr 2.9 cm 100 * LS / Open for AOI centered at X = 2.9 cm
AA R Corr 4.9 cm 100 * LS / Open for AOI centered at X = 4.9 cm
AB R Corr 6.9 cm 100 * LS / Open for AOI centered at X = 6.9 cm
AC Diff(X) -5.1 cm R corr minus avg R corr for AOI centered at X = -5.1 cm
AD Diff(X) -3.1 cm R corr minus avg R corr for AOI centered at X = -3.1 cm
AE Diff(X) -1.1 cm R corr minus avg R corr for AOI centered at X = -1.1 cm
AF Diff(X) 0.9 cm R corr minus avg R corr for AOI centered at X = 0.9 cm
AG Diff(X) 2.9 cm R corr minus avg R corr for AOI centered at X = 2.9 cm
AH Diff(X) 4.9 cm R corr minus avg R corr for AOI centered at X = 4.9 cm
AI Diff(X) 6.9 cm R corr minus avg R corr for AOI centered at X = 6.9 cm
AJ Avg of abs Diff(X) Average of absolute deviations (Diff Abs)
AK Machine Type Type of machine
AL Collimator Type of collimator
AM EPID Type of EPID
AN URL Link to main report.
AO outputPK Internal database reference to synthetic public key of output.
AP inputPK Internal database reference to synthetic public key of input.
AQ Uploaded By Anonymized version of user ID.
AR Pixel Rows Number of rows in the images. DICOM 0028,0010
AS Pixel Columns Number of columns in the images. DICOM 0028,0011
AT Pixel size X Image Plane Pixel Spacing. DICOM 3002,0011
AU Pixel size Y Image Plane Pixel Spacing. DICOM 3002,0011
AV b value for converting pixel values to CU: CU = m*PixelValue + b Rescale Intercept. DICOM 0028,1052
AW m value for converting pixel values to CU: CU = m*PixelValue + b Rescale Slope. DICOM 0028,1053
AX KVP Peak kilo voltage output of the X-Ray generator used. DICOM 0018,0060
AY Exposure Time msec Exposure time in mseync. DICOM 0016,0004
AZ Meterset Exposure Treatment machine Meterset duration over which image has been acquired, specified in Monitor units (MU) or minutes as defined by Primary Dosimeter Unit (300A,00B3). Required if Value 3 of Image Type (0008,0008) is PORTAL DICOM 3002,0032
BA Start Cumulative Meterset Wt Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition starts. DICOM 300C,0008
BB X-Ray Origin X Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
BC X-Ray Origin Y Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
BD X-Ray Origin Z Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
BE Source to Image Distance from radiation machine source to image plane (in mm) along radiation beam axis. DICOM 3002,0026
BF Source to Isocenter Radiation source to Gantry rotation axis distance of radiation machine used in acquiring or computing image (mm). DICOM 3002,0022
BG Gantry Angle Treatment machine gantry angle, i.e., orientation of IEC GANTRY coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees). DICOM 300A,011E
BH Collimator Angle Treatment machine beam limiting device (collimator) angle, i.e., orientation of IEC BEAM LIMITING DEVICE coordinate system with respect to IEC GANTRY coordinate system (degrees). DICOM 300A,0120
BI End Cumulative Meterset Wt Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition ends. DICOM 300C,0009
BJ Beam Number Uniquely identifies the corresponding N-segment treatment beam specified by Beam Number (300A,00C0) within Beam Sequence (300A,00B0) in RT Beams Module within the RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or the Ion Beam Sequence (300A,03A2) in the RT Ion Beams Module within the RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). DICOM 300C,0006
BK Operator Anonymized name(s) of the operator(s) supporting the Series. DICOM 0008,1070
BL Software Version Manufacturer's designation of software version of the equipment that produced the sources. DICOM 0018,1020
BM SOPInstanceUID Uniquely identifies the SOP Instance. DICOM 0008,0018
BN SeriesInstanceUID Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object instance(s). DICOM 0020,000E
BO PatientID Primary identifier for the Patient. DICOM 0010,0020
BP PatientName Patient's full name DICOM 0010,0010
BQ Open Pixel Rows Open Number of rows in the images. DICOM 0028,0010
BR Open Pixel Columns Open Number of columns in the images. DICOM 0028,0011
BS Open Pixel size X Open Image Plane Pixel Spacing. DICOM 3002,0011
BT Open Pixel size Y Open Image Plane Pixel Spacing. DICOM 3002,0011
BU Open b value for converting pixel values to CU: CU = m*PixelValue + b Open Rescale Intercept. DICOM 0028,1052
BV Open m value for converting pixel values to CU: CU = m*PixelValue + b Open Rescale Slope. DICOM 0028,1053
BW Open KVP Open Peak kilo voltage output of the X-Ray generator used. DICOM 0018,0060
BX Open Exposure Time msec Open Exposure time in mseync. DICOM 0016,0004
BY Open Meterset Exposure Open Treatment machine Meterset duration over which image has been acquired, specified in Monitor units (MU) or minutes as defined by Primary Dosimeter Unit (300A,00B3). Required if Value 3 of Image Type (0008,0008) is PORTAL DICOM 3002,0032
BZ Open Start Cumulative Meterset Wt Open Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition starts. DICOM 300C,0008
CA Open X-Ray Origin X Open Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
CB Open X-Ray Origin Y Open Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
CC Open X-Ray Origin Z Open Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
CD Open Source to Image Open Distance from radiation machine source to image plane (in mm) along radiation beam axis. DICOM 3002,0026
CE Open Source to Isocenter Open Radiation source to Gantry rotation axis distance of radiation machine used in acquiring or computing image (mm). DICOM 3002,0022
CF Open Gantry Angle Open Treatment machine gantry angle, i.e., orientation of IEC GANTRY coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees). DICOM 300A,011E
CG Open Collimator Angle Open Treatment machine beam limiting device (collimator) angle, i.e., orientation of IEC BEAM LIMITING DEVICE coordinate system with respect to IEC GANTRY coordinate system (degrees). DICOM 300A,0120
CH Open End Cumulative Meterset Wt Open Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition ends. DICOM 300C,0009
CI Open Beam Number Open Uniquely identifies the corresponding N-segment treatment beam specified by Beam Number (300A,00C0) within Beam Sequence (300A,00B0) in RT Beams Module within the RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or the Ion Beam Sequence (300A,03A2) in the RT Ion Beams Module within the RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). DICOM 300C,0006
CJ Open Operator Open Anonymized name(s) of the operator(s) supporting the Series. DICOM 0008,1070
CK Open Software Version Open Manufacturer's designation of software version of the equipment that produced the sources. DICOM 0018,1020
CL Open SOPInstanceUID Open Uniquely identifies the SOP Instance. DICOM 0008,0018
CM Open SeriesInstanceUID Open Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object instance(s). DICOM 0020,000E
CN Open PatientID Open Primary identifier for the Patient. DICOM 0010,0020
CO Open PatientName Open Patient's full name DICOM 0010,0010
CP Note User defined note for this output.