Column |
Column Title |
Description |
A |
Institution |
Anonymous version of institution name. |
B |
Machine |
Anonymous version of machine ID. |
C |
Acquisition |
Date and time that the first image in the data set was captured (delivery time). |
D |
Analysis |
Date and time that data was analyzed. |
E |
Procedure |
Procedure (test) used to process data |
F |
Beam Name |
Name of beam in plan |
G |
Gantry Angle Plan deg |
Planned gantry angle in degrees |
H |
gantryAnglePlan - Image deg |
Difference from planned gantry angle in degrees |
I |
collimatorAnglePlan deg |
Planned collimator angle in degrees |
J |
CollimatorAnglePlan - Image deg |
Difference from planned collimator angle in degrees |
K |
X1 Jaw Plan mm |
Planned jaw position for X1 in mm |
L |
X1 Jaw Plan - Image mm |
Difference from planned jaw position for X1 in mm |
M |
X2 Jaw Plan mm |
Planned jaw position for X2 in mm |
N |
X2 Jaw Plan - Image mm |
Difference from planned jaw position for X2 in mm |
O |
Y1 Jaw Plan mm |
Planned jaw position for Y1 in mm |
P |
Y1 Jaw Plan - Image mm |
Difference from planned jaw position for Y1 in mm |
Q |
Y2 Jaw Plan mm |
Planned jaw position for Y2 in mm |
R |
Y2 Jaw Plan - Image mm |
Difference from planned jaw position for Y2 in mm |
S |
Energy Plan KEV |
Planned energy in kilo electron volts |
T |
Energy Plan - Image KEV |
Difference from planned energy in kilo electron volts |
U |
Flattening Filter |
True if a flattening filter was present |
V |
Pass |
True if all values were within tolerances |
W |
Machine Type |
Type of machine |
X |
Collimator |
Type of collimator |
Y |
EPID |
Type of EPID |
Z |
URL |
Link to main report. |
AA |
outputPK |
Internal database reference to synthetic public key of output. |
AB |
inputPK |
Internal database reference to synthetic public key of input. |
AC |
Uploaded By |
Anonymized version of user ID. |
AD |
Pixel Rows |
Number of rows in the images. DICOM 0028,0010 |
AE |
Pixel Columns |
Number of columns in the images. DICOM 0028,0011 |
AF |
Pixel size X |
Image Plane Pixel Spacing. DICOM 3002,0011 |
AG |
Pixel size Y |
Image Plane Pixel Spacing. DICOM 3002,0011 |
AH |
b value for converting pixel values to CU: CU = m*PixelValue + b |
Rescale Intercept. DICOM 0028,1052 |
AI |
m value for converting pixel values to CU: CU = m*PixelValue + b |
Rescale Slope. DICOM 0028,1053 |
AJ |
KVP |
Peak kilo voltage output of the X-Ray generator used. DICOM 0018,0060 |
AK |
Exposure Time msec |
Exposure time in mseync. DICOM 0016,0004 |
AL |
Meterset Exposure |
Treatment machine Meterset duration over which image has been acquired, specified in Monitor units (MU) or minutes as defined by Primary Dosimeter Unit (300A,00B3). Required if Value 3 of Image Type (0008,0008) is PORTAL DICOM 3002,0032 |
AM |
Start Cumulative Meterset Wt |
Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition starts. DICOM 300C,0008 |
AN |
X-Ray Origin X |
Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D |
AO |
X-Ray Origin Y |
Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D |
AP |
X-Ray Origin Z |
Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D |
AQ |
Source to Image |
Distance from radiation machine source to image plane (in mm) along radiation beam axis. DICOM 3002,0026 |
AR |
Source to Isocenter |
Radiation source to Gantry rotation axis distance of radiation machine used in acquiring or computing image (mm). DICOM 3002,0022 |
AS |
Gantry Angle |
Treatment machine gantry angle, i.e., orientation of IEC GANTRY coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees). DICOM 300A,011E |
AT |
Collimator Angle |
Treatment machine beam limiting device (collimator) angle, i.e., orientation of IEC BEAM LIMITING DEVICE coordinate system with respect to IEC GANTRY coordinate system (degrees). DICOM 300A,0120 |
AU |
End Cumulative Meterset Wt |
Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition ends. DICOM 300C,0009 |
AV |
Beam Number |
Uniquely identifies the corresponding N-segment treatment beam specified by Beam Number (300A,00C0) within Beam Sequence (300A,00B0) in RT Beams Module within the RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or the Ion Beam Sequence (300A,03A2) in the RT Ion Beams Module within the RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). DICOM 300C,0006 |
AW |
Operator |
Anonymized name(s) of the operator(s) supporting the Series. DICOM 0008,1070 |
AX |
Software Version |
Manufacturer's designation of software version of the equipment that produced the sources. DICOM 0018,1020 |
AY |
SOPInstanceUID |
Uniquely identifies the SOP Instance. DICOM 0008,0018 |
AZ |
SeriesInstanceUID |
Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object instance(s). DICOM 0020,000E |
BA |
PatientID |
Primary identifier for the Patient. DICOM 0010,0020 |
BB |
PatientName |
Patient's full name DICOM 0010,0010 |
BC |
Note |
User defined note for this output. |