Definitions for GapOffsetSkew CSV columns.

Column Column Title Description
A Institution Anonymous version of institution name.
B Machine Anonymous version of machine ID.
C Acquisition Date and time that the first image in the data set was captured (delivery time).
D Analysis Date and time that data was analyzed.
E Procedure Procedure (test) used to process data
F C90 Gap (mm) Bank A avg - Bank B avg : Difference of averages between bank A and B
G C270 Gap (mm) Bank A avg - Bank B avg : Difference of averages between bank A and B
H C90 Offset (mm) (Bank A avg + Bank B avg) / 2 : Average of averages for bank A and B
I C270 Offset (mm) (Bank A avg + Bank B avg) / 2 : Average of averages for bank A and B
J C90 Bank A Skew (mm/40cm) (Bank A Right - Left / Horizontal Separation) * 400
K C270 Bank A Skew (mm/40cm) (Bank A Right - Left / Horizontal Separation) * 400
L C90 Bank B Skew (mm/40cm) (Bank B Right - Left / Horizontal Separation) * 400
M C270 Bank B Skew (mm/40cm) (Bank B Right - Left / Horizontal Separation) * 400
N C90 Bank A Right (mm) Vertical measurement of Bank A (X2) MLC leaves on the right-hand end.
O C270 Bank A Right (mm) Vertical measurement of Bank A (X2) MLC leaves on the right-hand end.
P C90 Bank A Left (mm) Vertical measurement of Bank A (X2) MLC leaves on the left-hand end.
Q C270 Bank A Left (mm) Vertical measurement of Bank A (X2) MLC leaves on the left-hand end.
R C90 Bank B Right (mm) Vertical measurement of Bank B (X1) MLC leaves on the right-hand end for
S C270 Bank B Right (mm) Vertical measurement of Bank B (X1) MLC leaves on the right-hand end for
T C90 Bank B Left (mm) Vertical measurement of Bank B (X1) MLC leaves on the left-hand end for
U C270 Bank B Left (mm) Vertical measurement of Bank B (X1) MLC leaves on the left-hand end for
V C90 Horizontal Separation (mm) Horizontal (x axis) distance between the centers of the right and left measurements. Used to calculate skew angle.
W C270 Horizontal Separation (mm) Horizontal (x axis) distance between the centers of the right and left measurements. Used to calculate skew angle.
X C90 Bank A Right - Left (mm) Bank A Right - Bank A Left showing the vertical change
Y C270 Bank A Right - Left (mm) Bank A Right - Bank A Left showing the vertical change
Z C90 Bank A Skew angle (deg) atan(Bank A Right - Left / Horizontal Separation)
AA C270 Bank A Skew angle (deg) atan(Bank A Right - Left / Horizontal Separation)
AB C90 Bank A avg (mm) (Bank A Right + Bank A Left) / 2 : Average of Bank A giving the vertical position of the midpoint
AC C270 Bank A avg (mm) (Bank A Right + Bank A Left) / 2 : Average of Bank A giving the vertical position of the midpoint
AD C90 Bank B Right - Left (mm) Bank B Right - Bank B Left showing the vertical change
AE C270 Bank B Right - Left (mm) Bank B Right - Bank B Left showing the vertical change
AF C90 Bank B Skew angle (deg) atan(Bank B Right - Left / Horizontal Separation)
AG C270 Bank B Skew angle (deg) atan(Bank B Right - Left / Horizontal Separation)
AH C90 Bank B avg (mm) (Bank B Right + Bank B Left) / 2 : Average of Bank B giving the vertical position of the midpoint
AI C270 Bank B avg (mm) (Bank B Right + Bank B Left) / 2 : Average of Bank B giving the vertical position of the midpoint
AJ C90 Right Diff (mm) Bank A Right - Bank B Right : Difference for each bank's right-hand values
AK C270 Right Diff (mm) Bank A Right - Bank B Right : Difference for each bank's right-hand values
AL C90 Left Diff (mm) Bank A Left - Bank B Left : Difference for each bank's left-hand values
AM C270 Left Diff (mm) Bank A Left - Bank B Left : Difference for each bank's left-hand values
AN C90 Diff of Diffs (mm) Bank A Right - Left - Bank B Right - Left : Difference for each bank's right and left-hand difference values
AO C270 Diff of Diffs (mm) Bank A Right - Left - Bank B Right - Left : Difference for each bank's right and left-hand difference values
AP C90 Skew Diff (deg) Bank A Skew angle - Bank B Skew angle : Difference of skew between bank A and B
AQ C270 Skew Diff (deg) Bank A Skew angle - Bank B Skew angle : Difference of skew between bank A and B
AR C90 Skew Diff (mm/40cm) Bank A Skew - Bank B Skew : Difference of skew between bank A and B
AS C270 Skew Diff (mm/40cm) Bank A Skew - Bank B Skew : Difference of skew between bank A and B
AT C90 Right Avg (mm) (Bank A Right + Bank B Right) / 2 : Average of each bank's right-hand values
AU C270 Right Avg (mm) (Bank A Right + Bank B Right) / 2 : Average of each bank's right-hand values
AV C90 Left Avg (mm) (Bank A Left + Bank B Left) / 2 : Average of each bank's left-hand values
AW C270 Left Avg (mm) (Bank A Left + Bank B Left) / 2 : Average of each bank's left-hand values
AX C90 Diff of A,B Diffs (mm) (Bank A Right - Left + Bank B Right - Left) / 2 : Average of each bank's right and left-hand differences
AY C270 Diff of A,B Diffs (mm) (Bank A Right - Left + Bank B Right - Left) / 2 : Average of each bank's right and left-hand differences
AZ C90 A,B Skew Angle Avg (deg) (Bank A Skew angle + Bank B Skew angle) / 2 : Average of skews for bank A and B
BA C270 A,B Skew Angle Avg (deg) (Bank A Skew angle + Bank B Skew angle) / 2 : Average of skews for bank A and B
BB C90 A,B Skew Avg (mm/40cm) (Bank A Skew + Bank B Skew) / 2 : Average of skews for bank A and B
BC C270 A,B Skew Avg (mm/40cm) (Bank A Skew + Bank B Skew) / 2 : Average of skews for bank A and B
BD J270 Gap (mm) Bank A avg - Jaw Avg : Difference of averages between bank A and Jaw
BE J270 Offset (mm) (Bank A avg + Jaw Avg) / 2 : Average of averages for bank A and Jaw
BF J270 Jaw Skew (mm/40cm) (Jaw Diff / Horizontal Separation) * 400
BG J270 Jaw Skew angle (deg) atan(Jaw Diff / Horizontal Separation)
BH J270 Jaw Right (mm) Measurement taken at right of edge of jaw near vertical center of image.
BI J270 Jaw Left (mm) Measurement taken at left of edge of jaw near vertical center of image.
BJ J270 Jaw Diff (mm) Jaw difference between Jaw Right - Jaw Left.
BK J270 Jaw Avg (mm) Jaw average (midpoint) between Jaw Right and Jaw Left.
BL J270 Right Diff (mm) Bank A Right - Jaw Right : Difference between bank A and jaw right-hand values
BM J270 Left Diff (mm) Bank A Left - Jaw Left : Difference between bank A and left-hand values
BN J270 Right-Left Diffs Diff (mm) Bank A Right - Left - Jaw Diff : Difference between bank A and jaw right-hand differences
BO J270 Skew Diff (mm/40cm) Bank A Skew - Jaw Skew : Difference of skew between bank A and Jaw
BP J270 Skew Angle Diff (deg) Bank A Skew angle - Jaw Skew angle : Difference of skew between bank A and Jaw
BQ J270 Right Avg (mm) (Bank A Right + Jaw Right) / 2 : Average of bank A and jaw right-hand values
BR J270 Left Avg (mm) (Bank A Left + Jaw Left) / 2 : Average of bank A and jaw left-hand values
BS J270 Right,Left Diffs Avg (mm) (Bank A Right - Left + Jaw Diff) / 2 : Average of bank A and jaw right and left-hand difference values
BT J270 Avg Skew (mm/40cm) (Bank A Skew + Jaw Skew) / 2 : Average of skews for bank A and Jaw
BU J270 Avg Skew Angle (deg) (Bank A Skew angle + Jaw Skew angle) / 2 : Average of skews for bank A and Jaw
BV ABank C90 Top Right Collimator 90 Bank A Y values measured at top right.
BW BBank C90 Top Right Collimator 90 Bank B Y values measured at top right.
BX ABank C270 Top Right Collimator 270 Bank A Y values measured at top right.
BY BBank C270 Top Right Collimator 270 Bank B Y values measured at top right.
BZ X1Jaw C270 Top Right Jaw 270 X1Y values measured at top right.
CA ABank C90 Top Left Collimator 90 Bank A Y values measured at top left.
CB BBank C90 Top Left Collimator 90 Bank B Y values measured at top left.
CC ABank C270 Top Left Collimator 270 Bank A Y values measured at top left.
CD BBank C270 Top Left Collimator 270 Bank B Y values measured at top left.
CE X1Jaw C270 Top Left Jaw 270 X1Y values measured at top left.
CF ABank C90 Bottom Right Collimator 90 Bank A Y values measured at bottom right.
CG BBank C90 Bottom Right Collimator 90 Bank B Y values measured at bottom right.
CH ABank C270 Bottom Right Collimator 270 Bank A Y values measured at bottom right.
CI BBank C270 Bottom Right Collimator 270 Bank B Y values measured at bottom right.
CJ X1Jaw C270 Bottom Right Jaw 270 X1Y values measured at bottom right.
CK ABank C90 Bottom Left Collimator 90 Bank A Y values measured at bottom left.
CL BBank C90 Bottom Left Collimator 90 Bank B Y values measured at bottom left.
CM ABank C270 Bottom Left Collimator 270 Bank A Y values measured at bottom left.
CN BBank C270 Bottom Left Collimator 270 Bank B Y values measured at bottom left.
CO X1Jaw C270 Bottom Left Jaw 270 X1Y values measured at bottom left.
CP Machine Type Type of machine
CQ Collimator Type of collimator
CR EPID Type of EPID
CS URL Link to main report.
CT outputPK Internal database reference to synthetic public key of output.
CU inputPK Internal database reference to synthetic public key of input.
CV Uploaded By Anonymized version of user ID.
CW ABank C90 Pixel Rows ABank C90 Number of rows in the images. DICOM 0028,0010
CX ABank C90 Pixel Columns ABank C90 Number of columns in the images. DICOM 0028,0011
CY ABank C90 Pixel size X ABank C90 Image Plane Pixel Spacing. DICOM 3002,0011
CZ ABank C90 Pixel size Y ABank C90 Image Plane Pixel Spacing. DICOM 3002,0011
DA ABank C90 b value for converting pixel values to CU: CU = m*PixelValue + b ABank C90 Rescale Intercept. DICOM 0028,1052
DB ABank C90 m value for converting pixel values to CU: CU = m*PixelValue + b ABank C90 Rescale Slope. DICOM 0028,1053
DC ABank C90 KVP ABank C90 Peak kilo voltage output of the X-Ray generator used. DICOM 0018,0060
DD ABank C90 Exposure Time msec ABank C90 Exposure time in mseync. DICOM 0016,0004
DE ABank C90 Meterset Exposure ABank C90 Treatment machine Meterset duration over which image has been acquired, specified in Monitor units (MU) or minutes as defined by Primary Dosimeter Unit (300A,00B3). Required if Value 3 of Image Type (0008,0008) is PORTAL DICOM 3002,0032
DF ABank C90 Start Cumulative Meterset Wt ABank C90 Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition starts. DICOM 300C,0008
DG ABank C90 X-Ray Origin X ABank C90 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
DH ABank C90 X-Ray Origin Y ABank C90 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
DI ABank C90 X-Ray Origin Z ABank C90 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
DJ ABank C90 Source to Image ABank C90 Distance from radiation machine source to image plane (in mm) along radiation beam axis. DICOM 3002,0026
DK ABank C90 Source to Isocenter ABank C90 Radiation source to Gantry rotation axis distance of radiation machine used in acquiring or computing image (mm). DICOM 3002,0022
DL ABank C90 Gantry Angle ABank C90 Treatment machine gantry angle, i.e., orientation of IEC GANTRY coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees). DICOM 300A,011E
DM ABank C90 Collimator Angle ABank C90 Treatment machine beam limiting device (collimator) angle, i.e., orientation of IEC BEAM LIMITING DEVICE coordinate system with respect to IEC GANTRY coordinate system (degrees). DICOM 300A,0120
DN ABank C90 End Cumulative Meterset Wt ABank C90 Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition ends. DICOM 300C,0009
DO ABank C90 Beam Number ABank C90 Uniquely identifies the corresponding N-segment treatment beam specified by Beam Number (300A,00C0) within Beam Sequence (300A,00B0) in RT Beams Module within the RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or the Ion Beam Sequence (300A,03A2) in the RT Ion Beams Module within the RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). DICOM 300C,0006
DP ABank C90 Operator ABank C90 Anonymized name(s) of the operator(s) supporting the Series. DICOM 0008,1070
DQ ABank C90 Software Version ABank C90 Manufacturer's designation of software version of the equipment that produced the sources. DICOM 0018,1020
DR ABank C90 SOPInstanceUID ABank C90 Uniquely identifies the SOP Instance. DICOM 0008,0018
DS ABank C90 SeriesInstanceUID ABank C90 Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object instance(s). DICOM 0020,000E
DT ABank C90 PatientID ABank C90 Primary identifier for the Patient. DICOM 0010,0020
DU ABank C90 PatientName ABank C90 Patient's full name DICOM 0010,0010
DV BBank C90 Pixel Rows BBank C90 Number of rows in the images. DICOM 0028,0010
DW BBank C90 Pixel Columns BBank C90 Number of columns in the images. DICOM 0028,0011
DX BBank C90 Pixel size X BBank C90 Image Plane Pixel Spacing. DICOM 3002,0011
DY BBank C90 Pixel size Y BBank C90 Image Plane Pixel Spacing. DICOM 3002,0011
DZ BBank C90 b value for converting pixel values to CU: CU = m*PixelValue + b BBank C90 Rescale Intercept. DICOM 0028,1052
EA BBank C90 m value for converting pixel values to CU: CU = m*PixelValue + b BBank C90 Rescale Slope. DICOM 0028,1053
EB BBank C90 KVP BBank C90 Peak kilo voltage output of the X-Ray generator used. DICOM 0018,0060
EC BBank C90 Exposure Time msec BBank C90 Exposure time in mseync. DICOM 0016,0004
ED BBank C90 Meterset Exposure BBank C90 Treatment machine Meterset duration over which image has been acquired, specified in Monitor units (MU) or minutes as defined by Primary Dosimeter Unit (300A,00B3). Required if Value 3 of Image Type (0008,0008) is PORTAL DICOM 3002,0032
EE BBank C90 Start Cumulative Meterset Wt BBank C90 Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition starts. DICOM 300C,0008
EF BBank C90 X-Ray Origin X BBank C90 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
EG BBank C90 X-Ray Origin Y BBank C90 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
EH BBank C90 X-Ray Origin Z BBank C90 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
EI BBank C90 Source to Image BBank C90 Distance from radiation machine source to image plane (in mm) along radiation beam axis. DICOM 3002,0026
EJ BBank C90 Source to Isocenter BBank C90 Radiation source to Gantry rotation axis distance of radiation machine used in acquiring or computing image (mm). DICOM 3002,0022
EK BBank C90 Gantry Angle BBank C90 Treatment machine gantry angle, i.e., orientation of IEC GANTRY coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees). DICOM 300A,011E
EL BBank C90 Collimator Angle BBank C90 Treatment machine beam limiting device (collimator) angle, i.e., orientation of IEC BEAM LIMITING DEVICE coordinate system with respect to IEC GANTRY coordinate system (degrees). DICOM 300A,0120
EM BBank C90 End Cumulative Meterset Wt BBank C90 Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition ends. DICOM 300C,0009
EN BBank C90 Beam Number BBank C90 Uniquely identifies the corresponding N-segment treatment beam specified by Beam Number (300A,00C0) within Beam Sequence (300A,00B0) in RT Beams Module within the RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or the Ion Beam Sequence (300A,03A2) in the RT Ion Beams Module within the RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). DICOM 300C,0006
EO BBank C90 Operator BBank C90 Anonymized name(s) of the operator(s) supporting the Series. DICOM 0008,1070
EP BBank C90 Software Version BBank C90 Manufacturer's designation of software version of the equipment that produced the sources. DICOM 0018,1020
EQ BBank C90 SOPInstanceUID BBank C90 Uniquely identifies the SOP Instance. DICOM 0008,0018
ER BBank C90 SeriesInstanceUID BBank C90 Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object instance(s). DICOM 0020,000E
ES BBank C90 PatientID BBank C90 Primary identifier for the Patient. DICOM 0010,0020
ET BBank C90 PatientName BBank C90 Patient's full name DICOM 0010,0010
EU ABank C270 Pixel Rows ABank C270 Number of rows in the images. DICOM 0028,0010
EV ABank C270 Pixel Columns ABank C270 Number of columns in the images. DICOM 0028,0011
EW ABank C270 Pixel size X ABank C270 Image Plane Pixel Spacing. DICOM 3002,0011
EX ABank C270 Pixel size Y ABank C270 Image Plane Pixel Spacing. DICOM 3002,0011
EY ABank C270 b value for converting pixel values to CU: CU = m*PixelValue + b ABank C270 Rescale Intercept. DICOM 0028,1052
EZ ABank C270 m value for converting pixel values to CU: CU = m*PixelValue + b ABank C270 Rescale Slope. DICOM 0028,1053
FA ABank C270 KVP ABank C270 Peak kilo voltage output of the X-Ray generator used. DICOM 0018,0060
FB ABank C270 Exposure Time msec ABank C270 Exposure time in mseync. DICOM 0016,0004
FC ABank C270 Meterset Exposure ABank C270 Treatment machine Meterset duration over which image has been acquired, specified in Monitor units (MU) or minutes as defined by Primary Dosimeter Unit (300A,00B3). Required if Value 3 of Image Type (0008,0008) is PORTAL DICOM 3002,0032
FD ABank C270 Start Cumulative Meterset Wt ABank C270 Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition starts. DICOM 300C,0008
FE ABank C270 X-Ray Origin X ABank C270 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
FF ABank C270 X-Ray Origin Y ABank C270 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
FG ABank C270 X-Ray Origin Z ABank C270 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
FH ABank C270 Source to Image ABank C270 Distance from radiation machine source to image plane (in mm) along radiation beam axis. DICOM 3002,0026
FI ABank C270 Source to Isocenter ABank C270 Radiation source to Gantry rotation axis distance of radiation machine used in acquiring or computing image (mm). DICOM 3002,0022
FJ ABank C270 Gantry Angle ABank C270 Treatment machine gantry angle, i.e., orientation of IEC GANTRY coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees). DICOM 300A,011E
FK ABank C270 Collimator Angle ABank C270 Treatment machine beam limiting device (collimator) angle, i.e., orientation of IEC BEAM LIMITING DEVICE coordinate system with respect to IEC GANTRY coordinate system (degrees). DICOM 300A,0120
FL ABank C270 End Cumulative Meterset Wt ABank C270 Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition ends. DICOM 300C,0009
FM ABank C270 Beam Number ABank C270 Uniquely identifies the corresponding N-segment treatment beam specified by Beam Number (300A,00C0) within Beam Sequence (300A,00B0) in RT Beams Module within the RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or the Ion Beam Sequence (300A,03A2) in the RT Ion Beams Module within the RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). DICOM 300C,0006
FN ABank C270 Operator ABank C270 Anonymized name(s) of the operator(s) supporting the Series. DICOM 0008,1070
FO ABank C270 Software Version ABank C270 Manufacturer's designation of software version of the equipment that produced the sources. DICOM 0018,1020
FP ABank C270 SOPInstanceUID ABank C270 Uniquely identifies the SOP Instance. DICOM 0008,0018
FQ ABank C270 SeriesInstanceUID ABank C270 Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object instance(s). DICOM 0020,000E
FR ABank C270 PatientID ABank C270 Primary identifier for the Patient. DICOM 0010,0020
FS ABank C270 PatientName ABank C270 Patient's full name DICOM 0010,0010
FT BBank C270 Pixel Rows BBank C270 Number of rows in the images. DICOM 0028,0010
FU BBank C270 Pixel Columns BBank C270 Number of columns in the images. DICOM 0028,0011
FV BBank C270 Pixel size X BBank C270 Image Plane Pixel Spacing. DICOM 3002,0011
FW BBank C270 Pixel size Y BBank C270 Image Plane Pixel Spacing. DICOM 3002,0011
FX BBank C270 b value for converting pixel values to CU: CU = m*PixelValue + b BBank C270 Rescale Intercept. DICOM 0028,1052
FY BBank C270 m value for converting pixel values to CU: CU = m*PixelValue + b BBank C270 Rescale Slope. DICOM 0028,1053
FZ BBank C270 KVP BBank C270 Peak kilo voltage output of the X-Ray generator used. DICOM 0018,0060
GA BBank C270 Exposure Time msec BBank C270 Exposure time in mseync. DICOM 0016,0004
GB BBank C270 Meterset Exposure BBank C270 Treatment machine Meterset duration over which image has been acquired, specified in Monitor units (MU) or minutes as defined by Primary Dosimeter Unit (300A,00B3). Required if Value 3 of Image Type (0008,0008) is PORTAL DICOM 3002,0032
GC BBank C270 Start Cumulative Meterset Wt BBank C270 Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition starts. DICOM 300C,0008
GD BBank C270 X-Ray Origin X BBank C270 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
GE BBank C270 X-Ray Origin Y BBank C270 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
GF BBank C270 X-Ray Origin Z BBank C270 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
GG BBank C270 Source to Image BBank C270 Distance from radiation machine source to image plane (in mm) along radiation beam axis. DICOM 3002,0026
GH BBank C270 Source to Isocenter BBank C270 Radiation source to Gantry rotation axis distance of radiation machine used in acquiring or computing image (mm). DICOM 3002,0022
GI BBank C270 Gantry Angle BBank C270 Treatment machine gantry angle, i.e., orientation of IEC GANTRY coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees). DICOM 300A,011E
GJ BBank C270 Collimator Angle BBank C270 Treatment machine beam limiting device (collimator) angle, i.e., orientation of IEC BEAM LIMITING DEVICE coordinate system with respect to IEC GANTRY coordinate system (degrees). DICOM 300A,0120
GK BBank C270 End Cumulative Meterset Wt BBank C270 Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition ends. DICOM 300C,0009
GL BBank C270 Beam Number BBank C270 Uniquely identifies the corresponding N-segment treatment beam specified by Beam Number (300A,00C0) within Beam Sequence (300A,00B0) in RT Beams Module within the RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or the Ion Beam Sequence (300A,03A2) in the RT Ion Beams Module within the RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). DICOM 300C,0006
GM BBank C270 Operator BBank C270 Anonymized name(s) of the operator(s) supporting the Series. DICOM 0008,1070
GN BBank C270 Software Version BBank C270 Manufacturer's designation of software version of the equipment that produced the sources. DICOM 0018,1020
GO BBank C270 SOPInstanceUID BBank C270 Uniquely identifies the SOP Instance. DICOM 0008,0018
GP BBank C270 SeriesInstanceUID BBank C270 Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object instance(s). DICOM 0020,000E
GQ BBank C270 PatientID BBank C270 Primary identifier for the Patient. DICOM 0010,0020
GR BBank C270 PatientName BBank C270 Patient's full name DICOM 0010,0010
GS Jaw X1 C270 Pixel Rows Jaw X1 C270 Number of rows in the images. DICOM 0028,0010
GT Jaw X1 C270 Pixel Columns Jaw X1 C270 Number of columns in the images. DICOM 0028,0011
GU Jaw X1 C270 Pixel size X Jaw X1 C270 Image Plane Pixel Spacing. DICOM 3002,0011
GV Jaw X1 C270 Pixel size Y Jaw X1 C270 Image Plane Pixel Spacing. DICOM 3002,0011
GW Jaw X1 C270 b value for converting pixel values to CU: CU = m*PixelValue + b Jaw X1 C270 Rescale Intercept. DICOM 0028,1052
GX Jaw X1 C270 m value for converting pixel values to CU: CU = m*PixelValue + b Jaw X1 C270 Rescale Slope. DICOM 0028,1053
GY Jaw X1 C270 KVP Jaw X1 C270 Peak kilo voltage output of the X-Ray generator used. DICOM 0018,0060
GZ Jaw X1 C270 Exposure Time msec Jaw X1 C270 Exposure time in mseync. DICOM 0016,0004
HA Jaw X1 C270 Meterset Exposure Jaw X1 C270 Treatment machine Meterset duration over which image has been acquired, specified in Monitor units (MU) or minutes as defined by Primary Dosimeter Unit (300A,00B3). Required if Value 3 of Image Type (0008,0008) is PORTAL DICOM 3002,0032
HB Jaw X1 C270 Start Cumulative Meterset Wt Jaw X1 C270 Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition starts. DICOM 300C,0008
HC Jaw X1 C270 X-Ray Origin X Jaw X1 C270 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
HD Jaw X1 C270 X-Ray Origin Y Jaw X1 C270 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
HE Jaw X1 C270 X-Ray Origin Z Jaw X1 C270 Position in (x,y,z) coordinates of origin of IEC X-RAY IMAGE RECEPTOR System in the IEC GANTRY coordinate system (mm). See Note 2. DICOM 3002,000D
HF Jaw X1 C270 Source to Image Jaw X1 C270 Distance from radiation machine source to image plane (in mm) along radiation beam axis. DICOM 3002,0026
HG Jaw X1 C270 Source to Isocenter Jaw X1 C270 Radiation source to Gantry rotation axis distance of radiation machine used in acquiring or computing image (mm). DICOM 3002,0022
HH Jaw X1 C270 Gantry Angle Jaw X1 C270 Treatment machine gantry angle, i.e., orientation of IEC GANTRY coordinate system with respect to IEC FIXED REFERENCE coordinate system (degrees). DICOM 300A,011E
HI Jaw X1 C270 Collimator Angle Jaw X1 C270 Treatment machine beam limiting device (collimator) angle, i.e., orientation of IEC BEAM LIMITING DEVICE coordinate system with respect to IEC GANTRY coordinate system (degrees). DICOM 300A,0120
HJ Jaw X1 C270 End Cumulative Meterset Wt Jaw X1 C270 Cumulative Meterset Weight within Beam referenced by Referenced Beam Number (300C,0006) at which image acquisition ends. DICOM 300C,0009
HK Jaw X1 C270 Beam Number Jaw X1 C270 Uniquely identifies the corresponding N-segment treatment beam specified by Beam Number (300A,00C0) within Beam Sequence (300A,00B0) in RT Beams Module within the RT Plan referenced in Referenced RT Plan Sequence (300C,0002) or the Ion Beam Sequence (300A,03A2) in the RT Ion Beams Module within the RT Ion Plan referenced in Referenced RT Plan Sequence (300C,0002). DICOM 300C,0006
HL Jaw X1 C270 Operator Jaw X1 C270 Anonymized name(s) of the operator(s) supporting the Series. DICOM 0008,1070
HM Jaw X1 C270 Software Version Jaw X1 C270 Manufacturer's designation of software version of the equipment that produced the sources. DICOM 0018,1020
HN Jaw X1 C270 SOPInstanceUID Jaw X1 C270 Uniquely identifies the SOP Instance. DICOM 0008,0018
HO Jaw X1 C270 SeriesInstanceUID Jaw X1 C270 Unique identifier for the Series that is part of the Study identified in Study Instance UID (0020,000D), if present, and contains the referenced object instance(s). DICOM 0020,000E
HP Jaw X1 C270 PatientID Jaw X1 C270 Primary identifier for the Patient. DICOM 0010,0020
HQ Jaw X1 C270 PatientName Jaw X1 C270 Patient's full name DICOM 0010,0010
HR Note User defined note for this output.